96-Probe Head
The Hamilton MICROLAB® STAR expands the high throughput of multichannel liquid handling with a 96 channel pipetting head.
The highly configurable and flexible Hamilton MICROLAB® STAR supports multiple arms, each with a multichannel pipetting head. Imagine the throughput and flexibility that can be achieved with both the 384 and 96 channel head on a single workstation with the highest on-deck plate density in the industry.
For re-usability and cost savings, the 96 head can use needles and disposable tips interchangeable on the fly without any change in hardware or service intervention.
The high precision 96 channel head is adaptable to a wide range of labware including 96, 384, and 1536 well microtiter plates.
Two channels on the he 96 channel head can detect liquid level conductively for dynamic vertical positioning and optimum tip position relative to the liquid level.
The 96 channel head has pipetting range of 1 – 1000μL. It excels in pipetting accuracy.
The 96 channel head can also jet dispense down to 1μL with touch off for applications requiring low-volume dispensing into dry microtiter plates.
Back to STAR Line
Specifications
Pipetting specifications |
| volume | precision | trueness |
for disposable tips | 10µl | 1µl | 5.0% | 5.0% |
10µl | 5µl | 2.0% | 2.5% | |
50µl | 5µl | 2.0% | 2.5% | |
50µl | 50µl | 1.0% | 1.5% | |
300µl | 50µl | 1.0% | 1.5% | |
1000µl | 1000µl | 1.0% | 1.0% |